Welcome to our online Request for Quote (RFQ) service, where you can fill out and submit the form provided below to obtain tailored procurement options from our team of industry experts. You have taken interest in part number 5790A-AF, which is a Voltage Standard item from Fluke Corporation that is listed with the NSN 6625-01-361-6809 for quick identification. Should you wish to move forward with quotation for this listing, be sure that you supply us with key details like your target pricing, quality, and delivery date. This information is what we will base presented quotes on, with responses offered shortly after the receipt and review of each form.
Note : We Will Not Share Your Information To Any Third Parties.
To ensure that customers only procure exact items of need, we present our offerings with as much relevant data as possible. For example, part number 5790A-AF was specifically produced by Fluke Corporation under the CAGE Code 89536, providing traceability. Furthermore, with the NSN 6625-01-361-6809, one can also determine that this listing falls under Federal Supply Class (FSC) 6625 Electrical and Electronic Properties Measuring and Testing Instruments. Alongside such details, all purchases will come delivered with any applicable qualifying certifications, furthering the reliability of our purchasing options. With this in mind, be sure to request a quote on part number 5790A-AF, and we will be in touch.
MRC | Criteria | Characteristic |
---|---|---|
BHXX | ELECTROMOTIVE FORCE PRODUCTION METHOD | ELECTRONIC |
BHYB | WARM-UP TIME FOR MAXIMUM ACCURACY IN MINUTES | 30.0 |
CRTL | CRITICALITY CODE JUSTIFICATION | SUPP |
AARB | TERMINAL TYPE | CONNECTOR AND BINDING POST |
APQB | UNIT TYPE | ENCASED |
ABMK | OVERALL WIDTH | 17.000 INCHES NOMINAL |
FAAZ | PHASE | SINGLE |
SR-5 | DESIGN CONTROL REFERENCE | 5790A/AF |
FEAT | SPECIAL FEATURES | FREQUENCY RANGE 10HZ TO 1MHZ; VOLTAGE RANGE 700UV TO 1000V; TRACEABLE TO NATIONAL STANDARDS; 24PPM TOTAL UNCERTAINTY; AUTOMATIC AC MEASUREMENT; COMPATIBLE WITH FLUKE A40 AND A40A CURRENT SHUNTS; POWER UP SELF TEST; 4 SETS OF INPUT TERMINALS,1 TYPE N AND 1 SET OF BINDING POSTS DEDICATED FOR AC MEASUREMENTS AND TRANSFER MODES. 2ND SET OF BINDING POST ARE FOR CURRENT SHUNTS;DIGITAL DISPLAY SHOWN IN PPM,PCT,VOLTS AND RATIO 1200V INPUT PROTECTION ON ALL RANGES:GPIB/IEEE488,RS-232C INTERFACES ARE INCLUDED; OPERATING TEMP 0 TO 50DEG C; OPERATING ALTITUDE 10000 FT; EMI/RFI COMPLIES FCC PART 15. SUBPART B,CLASS B,VDE 0871,CLASS B:ESD:EIA PN-1361,SURGE ANSI C62.41-1980 CAT ARELIABILITY: MIL-T-28800D PARA 3.13.3. WEIGHT: 53LBS; LINE VOLTAGES SELECTAB LE 100,110,115,120,200,220,230 OR 240VOLTS AC |
ABFY | OVERALL DEPTH | 24.800 INCHES NOMINAL |
SUPP | III SUPPLEMENTARY FEATURES | 5790A/AF IS CHECKED AND VERIFIED BY BY AEROSPACE GUIDENCE AND METROLOGY CENTER BEFORE BEING SHIPPED TO THE PMEL LABS |
BHXZ | OUTPUT ACCURACY RATING IN PERCENT | -0.010 TO 0.010 |
ACZB | FREQUENCY RATING | 47.0 HERTZ MINIMUM AND 63.0 HERTZ MAXIMUM |
SR-5 | THE MANUFACTURERS DATA | |
ACYN | AC VOLTAGE RATING | 115.0 VOLTS NOMINAL AND 220.0 VOLTS NOMINAL |
APCB | PORTABILITY | NONPORTABLE |
AARA | TERMINAL QUANTITY | 1 AND 5 |
ABKW | OVERALL HEIGHT | 7.000 INCHES NOMINAL |
SR-5 | MANUFACTURERS CODE | 89536 |